Philips Issues Worldwide Recall of Select Heartstart Fr2+ AED

Philips recently announced that it is voluntarily recalling approximately 5,400 HeartStart FR2+ automated external defibrillators (AEDs). This recall is being conducted due to the possibility of a memory chip failure that may render the device inoperable. Only certain HeartStart FR2+ AEDs (models M3860A and M3861A, distributed by Philips; and models M3840A and M3841A, distributed by Laerdal Medical) manufactured between May, 2007 and January, 2008 are included in the voluntary recall.

The HeartStart FR2+ defibrillators are used by trained responders and designated response teams to help treat sudden cardiac arrest. The device automatically analyzes the heart rhythm and determines whether a defibrillation shock is needed. If a shockable rhythm is detected, the FR2+ instructs the responder to deliver defibrillation therapy.

Philips has received reports of a memory chip failure in a small number of FR2+ units manufactured in 2007 and early 2008. These reported failures were detected during routine self tests, not during emergency use of the AED. Failure of this chip could render the AED inoperable and prevent it from delivering therapy when indicated, although Philips has received no reports of injury associated with this chip failure.

The AEDs affected by this recall have been distributed globally to fire departments, emergency medical services, hospitals, and other organizations. Philips is contacting customers to arrange for the return and replacement of all the recalled AEDs by sending notification letters to distributors and users. In addition, the company has set up a page on the Philips Web site with a serial number look-up tool to allow customers to find out if their FR2+ is part of this recall, as well as instructions on what to do if it is. The Web page is www.philips.com/FR2PlusAction.

Product Showcase

  • Full Line of Defense Against Combustible Dust Nilfisk

    Nilfisk provides a comprehensive range of industrial vacuums meticulously crafted to adhere to NFPA 652 housekeeping standards, essential for gathering combustible dust in Class I, Group D, and Class II, Groups E, F & G environments or non-classified settings. Our pneumatic vacuums are meticulously engineered to fulfill safety criteria for deployment in hazardous surroundings. Leveraging advanced filtration technology, Nilfisk ensures the secure capture of combustible materials scattered throughout your facility, ranging from fuels, solvents, and metal dust to flour, sugar, and pharmaceutical powders. Read More

  • HAZ LO HEADLAMPS

    With alkaline or rechargeable options, these safety rated, Class 1, Div. 1 Headlamps provide long runtime with both spot and flood options in the same light. Work safely and avoid trip hazards with flexible hands-free lighting from Streamlight. Read More

  • Preventative Heat Safety

    Dehydration and heat exposure impair physical and cognitive performance. Proper hydration boosts heat stress resilience, but hydration needs are highly individualized and hard to predict across a workforce. Connected Hydration® empowers industrial athletes to stay safe through behavioral interventions, informed by sports science, and equips safety teams with critical insights to anticipate high-risk situations and adapt to evolving environmental factors. Curious about applying the latest in sports science based hydration strategies for industrial athletes? Stop by booth #1112 at AIHA or schedule a free demo today at https://epcr.cc/demo. Read More

Featured

Artificial Intelligence

Webinars