FDA Posts Notice on Medtronic Cranial Software Recall

Medtronic recalled Synergy Cranial Software and Stealth Station S7 Cranial Software in September 2018 because of inaccuracies displayed during surgical procedures. The notice says FDA has identified this as a Class I recall, the most serious type of recall.

The U.S. Food and Drug Administration recently added a new MedWatch Safety Alert recall notice to the FDA Recalls page concerning Medtronic's recalled Synergy Cranial Software and Stealth Station S7 Cranial Software. They were recalled in September 2018 because of inaccuracies displayed during surgical procedures, according to the notice.

It says FDA has identified this as a Class I recall, the most serious type of recall, because use of the software may cause serious injuries or death. The recall affects model numbers 9733763 with software versions 2.2.0, 2.2.5, 2.2.6, 2.2.7; and 9735585 with software versions 3.0.0, 3.0.1, 3.0.2, 3.1.0. These were manufactured and distributed from Aug. 31, 2011, to Oct. 16, 2018. A total of 5,487 have been recalled, according to the notice.

The products are used with the StealthStation Surgical Navigation System to provide detailed three-dimensional images of a patient's brain to help neurosurgeons safely navigate surgical tools and implants used during brain surgery. Medtronic recalled them due to reports of incorrect information displaying during biopsy procedures that could result in serious or life-threatening patient harm; according to the notice, during a biopsy procedure, the software monitor may show that the tip of the surgical tool has not yet reached the planned target and may prevent the neurosurgeon from being able to accurately see the location of surgical tools in the patient's brain. In the event this software defect occurs, the neurosurgeon could potentially insert the surgical tool too deeply and damage the patient's healthy tissue, brain or blood vessels.

Medtronic on Sept. 21, 2018, sent an Urgent Medical Device Correction notice to customers. The company reported Dec. 28, 2018, that it was continuing to follow up with customers about the problem.

Product Showcase

  • EMSL Analytical, Inc. - Air Sampling Supplies & Testing Labs

    EMSL Analytical, Inc. operates laboratories throughout the United States and Canada. EMSL is a nationally recognized and locally focused provider specializing in fast laboratory results for Asbestos, Mold, Silica, Lead & Metals, Bacteria, Legionella, USP , Combustion By-Products, VOC’s, Radon, PCB’s, Formaldehyde, METH/Fentanyl, Identification of Dust & Unknowns. Sampling Pumps, Cassettes, Media & Supplies available. Reach us at 1-800.220.3675 Read More

  • Magid® D-ROC® GPD412 21G Ultra-Thin Polyurethane Palm Coated Work Gloves

    Magid’s 21G line is more than just a 21-gauge glove, it’s a revolutionary knitting technology paired with an advanced selection of innovative fibers to create the ultimate in lightweight cut protection. The latest offering in our 21G line provides ANSI A4 cut resistance with unparalleled dexterity and extreme comfort that no other 21-gauge glove on the market can offer! Read More

  • The MGC Simple Plus

    The MGC Simple Plus is a simple-to-use, portable multi gas detector that runs continuously for three years without being recharged or routinely calibrated after its initial charge and calibration during manufacturing. The detector reliably tests a worksite’s atmosphere for hydrogen sulfide, carbon monoxide, oxygen and combustible gases (LEL). Its durability enables the detector to withstand the harshest treatment and environments, hence earning it an IP 68 rating. The MGC Simple Plus is also compatible with a variety of accessories, such as the GCT External Pump. Visit gascliptech.com for more information. Read More

Featured

Artificial Intelligence